PUMP INFORMATION


Zyno Z-800F

The Zyno Z-800F infusion pump has four primary purpose-driven design objectives that are central to the development efforts: simplicity, saftey, reliability, and durability. The purpose-driven design objectives were met to create a diverse range of clinical use-cases necessary for the infusion pump environment. This provides a pump that is intelligent with multi-therapy programming modes that include: Continuous Rate / Time / Volume Infusion.

Unique Features: The Zyno pump mounts at a 45 degree angle for improved visibility and usability. There is a visual indicator for improved alarm identification and response. It has a removable power plug for improved portability.

Flow Rates: The Z -800F Infusion Pump flow rate range is from 1 to 1200 ml/h.

Volume-To-Be-Infused (VTBI): The volume-to-be-infused (VTBI) range is from 1 to 9999 ml in 1 ml increments.

Free Flow Protection: The Z-800F Infusion Pump has a built in free flow clamp to prevent inadvertent free flow when the set is loaded in the pump. The Z-800F Infusion Pump also has an IV Set based anti-free-flow module to prevent inadvertent free flow when the set is unloaded from the pump.

Occlusion Pressure: The Z-800F Infusion Pump provides 14 levels of adjustable downstream occlusion alarm thresholds between 4 psi and 30 psi.

Secondary Infusions: Secondary infusions may be delivered at a specified secondary delivery rate and secondary volume-to-be-infused (VTBI) independent of the primary infusion parameters. When secondary VTBI is greater than zero, the pump will always execute secondary infusion first. Automatic changeover occurs to the primary infusion parameters when the secondary infusion is complete. A Zyno Medical proprietary primary administration set with a back check valve must be used.

Battery Operating Time: 8 hours at 125 ml/hr.